CAUTION:
Indicates a potentially hazardous situation, if not avoided, could result in minor or
moderate injury and property damage.
• The MaxVenturi is intended for use with specific patient interface configurations. The numbered
scale is intended for use with the Fisher & Paykel* heated humidifier (MR850) and the Optiflow
high flow patient interface system (OPT544, OPT546, OPT570). The lettered scale (Labeled A
through F) is intended for the alternate patient circuits found on the chart provided in Section 2.
• Never install the sensor in a location that will expose the sensor to patient exhalation
or secretions.
• Only use Maxtec approved accessories and replacement parts. Failure to do so may seriously
impair the MaxVenturi’s performance. Repair or alteration of the MaxVenturi beyond the scope
of the maintenance instructions or by anyone other than authorized Maxtec service personnel
could cause the product to fail to perform as designed.
• Use of the MaxVenturi near devices that generate strong electrical fields may cause
erratic readings.
• This device has a visual low-battery alarm but no audible alarm.
• Calibrate the MaxVenturi weekly when in operation or if environmental conditions change
significantly, i.e., Temperature, Humidity, Barometric Pressure. (refer to Calibration sections of
this manual).
• The device will assume a percent oxygen concentration when calibrating. Be sure to apply
100% oxygen or room air concentration to the device during calibration or the device will not
calibrate correctly. (See section 2.2)
• Do not use ethylene oxide sterilization. Do not immerse the device in any cleaning solution,
autoclave or expose the device to high temperatures.
• Use with any other patient interface system may result in false readings from the
flow meter.
• The MaxVenturi is not intended for steam, ethylene oxide or radiation sterilization.
• Do not clean with ethanol or acetone.
• After cleaning, and before use on a patient, attach the device to an oxygen supply and allow the
device to run at high flow for several minutes to allow any cleaning fluids or vapors to evaporate
and be flushed out.
• The device can leak excess oxygen out the entrainment port if the oxygen control knob is
turned up too high. This can lead to a minor drop in total air flow to the patient and excess
oxygen entering the room environment.
• Federal Law (USA) restricts this device to sale by or on the order of a physician.
ATTENTION:
Indicates a potentially hazardous situation, which, if not avoided could result in
property damage.
• If the MaxVenturi is ever exposed to liquids (spills or immersion) or to any other physical abuse,
return to Maxtec for evaluation before use.
• Always remove the batteries to protect the unit from potential leaky battery damage when the
unit is going to be stored or not in use for periods exceeding 1 month. Replace dead batteries
with high quality AA Alkaline batteries.
• Do not use rechargeable batteries with this device.
• Maxtec cannot warranty any damage resulting from misuse, unauthorized repair or improper
maintenance of the instrument.
• This product is Latex free.
• Avoid usage in environments with greater than 95% relative humidity.